PWL-CX uses an imported PTFE single electrode and self-sealing structure to measure level from −200 to 250°C and up to 32 MPa without leakage-prone moving parts.
Applications
High-pressure equipment
Pressurized process vessels
High-temperature containers
Cryogenic tanks
Liquid CO2, liquid nitrogen and LNG
Sealed pressure tanks and vessels
Model | PWL-CX |
Probe Diameter | Φ8, Φ12, Φ14,Φ16, Φ19 |
Order Length | 50-10000mm |
Working Pressure | ≤0.6MPa, ≤32MPa (in normal temperature) |
Accuracy | ±1.0%FS(Typical), ±2.5%FS for Cryogenic |
Environmental Temp. | -40℃~+85℃ |
Temp. Resistance of Electrode | -60℃~+120℃,-200℃~+200℃, -200℃~+250℃ |
Electrical Connection | Fixed cable by stainless steel cap or industrial wire interface |
IP Rating | IP67, IP65 |
Cable Length | 1000mm or by customized |
Signal Output & Power Supply | 4-20mA 2 wires RS485 Modbus 18-28VDC 12-28VDC |
Pressure Connection | M20*1.5, M27*2 Flange connection by customization |
Long-term Stability (%FS/year) | ≤0.1 |
Temp Drift ( %FS/℃ ) | ≤0.02 |
Probe Material | SUS304, PTFE covered SUS304 |
Explosion Proof | Exia II CT6 Ga |